Single Pulse Laser Induced Breakdown Spectroscopy (SP-LIBS), performed by nanosecond and femtosecond laser sources has been applied to the study of the depth profiling of the artificially obtained patina of a bronze sample. The results show improved performances of femtosecond LIBS compared to nanosecond one. The differences found in the analyses are related to the different laser-matter interaction processes induced by the different time duration of the laser pulses. © 2013 Elsevier B.V.

Comparison of the performances of nanosecond and femtosecond Laser Induced Breakdown Spectroscopy for depth profiling of an artificially corroded bronze

Santagata A;
2014

Abstract

Single Pulse Laser Induced Breakdown Spectroscopy (SP-LIBS), performed by nanosecond and femtosecond laser sources has been applied to the study of the depth profiling of the artificially obtained patina of a bronze sample. The results show improved performances of femtosecond LIBS compared to nanosecond one. The differences found in the analyses are related to the different laser-matter interaction processes induced by the different time duration of the laser pulses. © 2013 Elsevier B.V.
2014
Istituto di Struttura della Materia - ISM - Sede Roma Tor Vergata
Inglese
302
275
279
http://www.scopus.com/inward/record.url?eid=2-s2.0-84897394146&partnerID=q2rCbXpz
Sì, ma tipo non specificato
Copper alloy
Depth profiling
fs laser
ns laser
Single-pulse LIBS
6
info:eu-repo/semantics/article
262
De Bonis, A; De Filippo, B; Galasso, A; Santagata, A; Smaldone, A; Teghil, R
01 Contributo su Rivista::01.01 Articolo in rivista
none
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/262625
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 20
  • ???jsp.display-item.citation.isi??? ND
social impact