The calibration-free methodology described in the literature dealing with the technique of laser induced breakdown spectroscopy has been applied here to several samples under different operating pressures with the aim of classifying the analytical technique as quantitative, semi-quantitative or qualitative (with quantitative estimates only), depending on the relative percentage errors calculated for the major, minor and trace elemental composition of the sample. The results were obtained with a Q-switched Nd:YAG laser (1064 nm), characterized by 5.3 ns duration pulses at a maximum repetition frequency of 10 Hz and with pulse energy up to 300 mJ. The spectra were taken with a grating monochromator coupled with an intensified CCD detector, calibrated from 220 to 700 nm. The samples investigated were aluminum alloy standard disks, standard reference materials of Zn alloy, Cu-Ni alloy, Ti alloy, Ni alloy powder, Cu-Zn alloy and soils. The data are compared with those published in the literature. The results show that the best accuracy (5% relative error) is usually obtained for the matrix element, while in most cases, for the remainder of the components in the sample (major, minor or trace composition), the results can only be considered semi-quantitative (<200% relative error).

Semi-quantitative analysis of metal alloys, brass and soil samples by calibration-free laser-induced breakdown spectroscopy: recent results and considerations

E Tognoni;
2009

Abstract

The calibration-free methodology described in the literature dealing with the technique of laser induced breakdown spectroscopy has been applied here to several samples under different operating pressures with the aim of classifying the analytical technique as quantitative, semi-quantitative or qualitative (with quantitative estimates only), depending on the relative percentage errors calculated for the major, minor and trace elemental composition of the sample. The results were obtained with a Q-switched Nd:YAG laser (1064 nm), characterized by 5.3 ns duration pulses at a maximum repetition frequency of 10 Hz and with pulse energy up to 300 mJ. The spectra were taken with a grating monochromator coupled with an intensified CCD detector, calibrated from 220 to 700 nm. The samples investigated were aluminum alloy standard disks, standard reference materials of Zn alloy, Cu-Ni alloy, Ti alloy, Ni alloy powder, Cu-Zn alloy and soils. The data are compared with those published in the literature. The results show that the best accuracy (5% relative error) is usually obtained for the matrix element, while in most cases, for the remainder of the components in the sample (major, minor or trace composition), the results can only be considered semi-quantitative (<200% relative error).
2009
Istituto per i Processi Chimico-Fisici - IPCF
File in questo prodotto:
File Dimensione Formato  
prod_179904-doc_16928.pdf

non disponibili

Descrizione: ARTICOLO PUBBLICATO
Dimensione 3.07 MB
Formato Adobe PDF
3.07 MB Adobe PDF   Visualizza/Apri   Richiedi una copia