An X-ray diffractometer for small-angle scattering and diffraction experiments using synchroton radiation is described. The diffractometer is in operation at the Adone storage ring at Frascati. The optical collimation is 8 × 10-5 rad. A three-dimensional goniometer carriers either a two-dimensional position-sensitive drift chamber detector or a one-dimensional cylindrical drift chamber. The detectors can move along a spherical surface with radius 500 mm within an arc of 0.35 rad. The best angular resolution of the diffractometer, determined by the best resolution of the detector, is 2 × 10-4 rad.
An X-ray diffractometer for synchrotron radiation
Allocca L;
1984
Abstract
An X-ray diffractometer for small-angle scattering and diffraction experiments using synchroton radiation is described. The diffractometer is in operation at the Adone storage ring at Frascati. The optical collimation is 8 × 10-5 rad. A three-dimensional goniometer carriers either a two-dimensional position-sensitive drift chamber detector or a one-dimensional cylindrical drift chamber. The detectors can move along a spherical surface with radius 500 mm within an arc of 0.35 rad. The best angular resolution of the diffractometer, determined by the best resolution of the detector, is 2 × 10-4 rad.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.