Single-crystal diamond Schottky photodiodes have been developed following a WC-diamond-TiC/Ti/Ag vertical structure for X-ray dosimetry with characteristics of low leakage current and the capability of zero-bias operations. Continuous and modulated X-ray measurements show that the devices developed provide a linear response to dose rate at low bias voltages (< 2.5 V) and at modulation frequencies > 130 Hz. This supports their application in modulated radiotherapy treatments.

Dosimetric Performance of Single-Crystal Diamond X-Ray Schottky Photodiodes

A Bellucci;S Orlando;E Cappelli;D M Trucchi
2013

Abstract

Single-crystal diamond Schottky photodiodes have been developed following a WC-diamond-TiC/Ti/Ag vertical structure for X-ray dosimetry with characteristics of low leakage current and the capability of zero-bias operations. Continuous and modulated X-ray measurements show that the devices developed provide a linear response to dose rate at low bias voltages (< 2.5 V) and at modulation frequencies > 130 Hz. This supports their application in modulated radiotherapy treatments.
2013
Istituto di Struttura della Materia - ISM - Sede Roma Tor Vergata
Istituto di Struttura della Materia - ISM - Sede Roma Tor Vergata
Diamond
fast response
linearity
radiation dosimeter
Schottky diode.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/15581
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 12
  • ???jsp.display-item.citation.isi??? ND
social impact