A capillary electrophoresis apparatus was used as sampler for flow injection analysis (FIA) in tandem mass spectrometry of l-carnitine and its acetyl- and propionyl-metabolites in human plasma. The capillary electrophoresis instrument was coupled to the ion trap mass spectrometer by an electrospray ionization coaxial sheath liquid interface. The electrophoresis capillary introduced the sample directly into the source by applying a prolonged sample injection. The use of the capillary electrophoresis apparatus miniaturised the FIA procedure, substantially reducing the quantities of solvents and samples used, and allowed rapid automated sequential analyses. The method was optimised and validated using a dialyzed human plasma matrix. The plasma samples were analysed after a simple, rapid deproteinisation procedure with acetonitrile and diluted 70 times before direct injection into the mass spectrometer for product ion scan MS/MS analysis in positive ionisation. The total analysis time was 5 min, including capillary preconditioning and acquisition time (3 min). The method was sensitive, allowing the determination of l-, l-acetyl- and l-propionyl-carnitines at 140, 14 and 3.6nM concentrations (injected values) corresponding to lower limit of quantitation values in plasma of 10, 1 and 0.25M, respectively. The method was processed for full validation and applied to the analysis of l-carnitine and its short chain derivatives in human plasma samples.

Rapid determination of short chain carnitines in human plasma by electrospray ionisation-ion trap mass spectrometry using capillary electrophoresis instrument as sampler

Claudia Desiderio;Massimo Castagnola
2007

Abstract

A capillary electrophoresis apparatus was used as sampler for flow injection analysis (FIA) in tandem mass spectrometry of l-carnitine and its acetyl- and propionyl-metabolites in human plasma. The capillary electrophoresis instrument was coupled to the ion trap mass spectrometer by an electrospray ionization coaxial sheath liquid interface. The electrophoresis capillary introduced the sample directly into the source by applying a prolonged sample injection. The use of the capillary electrophoresis apparatus miniaturised the FIA procedure, substantially reducing the quantities of solvents and samples used, and allowed rapid automated sequential analyses. The method was optimised and validated using a dialyzed human plasma matrix. The plasma samples were analysed after a simple, rapid deproteinisation procedure with acetonitrile and diluted 70 times before direct injection into the mass spectrometer for product ion scan MS/MS analysis in positive ionisation. The total analysis time was 5 min, including capillary preconditioning and acquisition time (3 min). The method was sensitive, allowing the determination of l-, l-acetyl- and l-propionyl-carnitines at 140, 14 and 3.6nM concentrations (injected values) corresponding to lower limit of quantitation values in plasma of 10, 1 and 0.25M, respectively. The method was processed for full validation and applied to the analysis of l-carnitine and its short chain derivatives in human plasma samples.
2007
Istituto di Chimica del Riconoscimento Molecolare - ICRM - Sede Milano
Flow injection analysis; Carnitine; Plasma; Miniaturization; Capillary electrophoresis
File in questo prodotto:
File Dimensione Formato  
prod_17499-doc_3825.pdf

non disponibili

Descrizione: pubblicazione
Dimensione 413.95 kB
Formato Adobe PDF
413.95 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/167480
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 9
social impact