Comparison measurements on reference standards are reported in which 13 partners with different instruments took part. A set of prototype standards which had been produced and calibrated within a European project were used for the measurements. Here, results of measurements on a 240 nm step height standard and a two-dimensional lateral standard with a nominal pitch of 1 m are reported.
Intercomparison of scanning probe microscopes
2002
Abstract
Comparison measurements on reference standards are reported in which 13 partners with different instruments took part. A set of prototype standards which had been produced and calibrated within a European project were used for the measurements. Here, results of measurements on a 240 nm step height standard and a two-dimensional lateral standard with a nominal pitch of 1 m are reported.File in questo prodotto: