Frequency dependent radial compliance information is used to determine the maximum effective test speed when performing accelerated durability testing on vascular graft samples. The modulus versus frequency data obtained from the experiment prove that the dynamic properties of the resultant device can be affected by changing the polyurethane vascular graft structure.

Relating the frequency dependent radial compliance to the tensile modulus of plyurethane vascular grafts

Soldani G;
1996

Abstract

Frequency dependent radial compliance information is used to determine the maximum effective test speed when performing accelerated durability testing on vascular graft samples. The modulus versus frequency data obtained from the experiment prove that the dynamic properties of the resultant device can be affected by changing the polyurethane vascular graft structure.
1996
Istituto di Fisiologia Clinica - IFC
Radial Compliance
Tensile modulus
Polyurethane Vascular grafts
File in questo prodotto:
File Dimensione Formato  
prod_240653-doc_117909.pdf

solo utenti autorizzati

Descrizione: Relating the Frequency Dependent Radial Compliance to the Tensile Modulus of Polyurethane Vascular Grafts
Tipologia: Versione Editoriale (PDF)
Dimensione 3.07 MB
Formato Adobe PDF
3.07 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
prod_240653-doc_117910.pdf

solo utenti autorizzati

Descrizione: Relating the Frequency Dependent Radial Compliance to the Tensile Modulus of Polyurethane Vascular Grafts
Tipologia: Versione Editoriale (PDF)
Dimensione 416.79 kB
Formato Adobe PDF
416.79 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/17122
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? ND
social impact