The system the intercomparison of unlike impedances of IEN, based on the three-volumeter method and four-terminal pair definition of the standards, has now been provided with a new direct-digital synthesis generator, having multiple outputs, isolation among the channels, and high amplitude and frequency stability. Experimental results in the new and old configuration are shown and compared.

Impedance measurement by means of a high-stability multiphase DDS generator, with the three-voltage method

Galzerano G;
2002

Abstract

The system the intercomparison of unlike impedances of IEN, based on the three-volumeter method and four-terminal pair definition of the standards, has now been provided with a new direct-digital synthesis generator, having multiple outputs, isolation among the channels, and high amplitude and frequency stability. Experimental results in the new and old configuration are shown and compared.
2002
Istituto di fotonica e nanotecnologie - IFN
Inglese
IMTC 2002: PROCEEDINGS OF THE 19TH IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1 & 2
19th IEEE Instrumentation and Measurement Technology Conference (IMTC/2002)
1183
1185
0-7803-7218-2
AC generators
impedance measurement
standards
electrical variables measurement
SINEWAVE GENERATOR
19th IEEE Instrumentation and Measurement Technology Conference (IMTC/2002), ANCHORAGE, AK, MAY 21-23, 2002
4
none
Callegaro, L; Galzerano, G; Svelto, ; C,
273
info:eu-repo/semantics/conferenceObject
04 Contributo in convegno::04.01 Contributo in Atti di convegno
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/18337
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 2
social impact