We have implemented a novel scheme of signal readout for resonant gravitational wave detectors. For the first time, a capacitive resonant transducer has been matched to the signal amplifier by means of a tuned high Q electrical resonator. The resulting 3-mode detection scheme widens significantly the bandwidth of the detector. We present here the results achieved by this signal readout equipped with a two-stage SQUID amplifier. Once installed on the AURIGA detector, the one-sided spectral sensitivity obtained with the detector operated at 4.5 K is better than 10(-20) Hz(-1/2) over 110 Hz, and in good agreement with the expectations.
3-Mode detection for widening the bandwidth of resonant gravitational wave detectors
Bonaldi M;Falferi P;Vinante A;
2005
Abstract
We have implemented a novel scheme of signal readout for resonant gravitational wave detectors. For the first time, a capacitive resonant transducer has been matched to the signal amplifier by means of a tuned high Q electrical resonator. The resulting 3-mode detection scheme widens significantly the bandwidth of the detector. We present here the results achieved by this signal readout equipped with a two-stage SQUID amplifier. Once installed on the AURIGA detector, the one-sided spectral sensitivity obtained with the detector operated at 4.5 K is better than 10(-20) Hz(-1/2) over 110 Hz, and in good agreement with the expectations.File in questo prodotto: