We investigated possible variations of impedance values in samples of sodium chloride solution (sodium chloride 0.9%) with glucose at different concentrations, ranging from 5000 to around 75 mg/dL. The sodium chloride solution (either saline physiological solution) was chosen since it has similarities to blood but no cell components, which may be confounding factors in this study. Special focus was on the effect of stirring and of temperature variations on the impedance spectrum of samples at different glucose concentrations. We found that variations in glucose concentration directly affect the impedance modulus of the sample both in static conditions and in dynamic conditions due to stirring, as well as at both room temperature and at increased temperature. In fact, even if the impedance variations were often small (around 3-4 m ? per mg/dL), they were usually clearly measurable. These findings may be the basis for possible development of a new approach, based on impedance technology, for the noninvasive monitoring of glycaemia.

Electroimpedance spectroscopy for the measurement of the dielectric properties of sodium chloride solutions at different glucose concentrations

Sbrignadello S;Tura A;Ravazzani P
2013

Abstract

We investigated possible variations of impedance values in samples of sodium chloride solution (sodium chloride 0.9%) with glucose at different concentrations, ranging from 5000 to around 75 mg/dL. The sodium chloride solution (either saline physiological solution) was chosen since it has similarities to blood but no cell components, which may be confounding factors in this study. Special focus was on the effect of stirring and of temperature variations on the impedance spectrum of samples at different glucose concentrations. We found that variations in glucose concentration directly affect the impedance modulus of the sample both in static conditions and in dynamic conditions due to stirring, as well as at both room temperature and at increased temperature. In fact, even if the impedance variations were often small (around 3-4 m ? per mg/dL), they were usually clearly measurable. These findings may be the basis for possible development of a new approach, based on impedance technology, for the noninvasive monitoring of glycaemia.
2013
Istituto di Elettronica e di Ingegneria dell'Informazione e delle Telecomunicazioni - IEIIT
INGEGNERIA BIOMEDICA
Istituto di Neuroscienze - IN -
Glucose concentration
Impedance spectrum
Impedance variations
Increased temperature
Non-invasive monitoring
Physiological solution
Sodium chloride solution
Temperature variation
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/218352
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 14
  • ???jsp.display-item.citation.isi??? ND
social impact