A fluorescence polarization immunoassay (FPIA) has been commercially released for routine large-scale testing of total homocysteine (tHcy) on the AxSYM analyzer. We evaluated the analytical performance of the AxSYM tHcy FPIA and compared it with the well established high-performance liquid chromatography (HPLC) and IMx tHcy FPIA methods. Homocysteine concentrations were measured by AxSYM and IMx tHcy FPIA and by a rapid isocratic HPLC method with fluorescence detection. Coefficient of variation (CV) of total imprecision for AxSYM tHcy was < or = 5%, mean dilution recovery 102%, analytical sensitivity 0.70 micromol/l and linearity was good up to 1:8 dilution. Spearman rank correlations, rho, were 0.83 (p < 0.0001) for AxSYM vs. HPLC, 0.97 (p < 0.0001) for AxSYM vs. IMx and 0.83 (p < 0.0001) for IMx vs. HPLC. Passing and Bablok regression Y-intercepts and slopes were: 2.944/0.937 (AxSYM vs. HPLC), -0.367/ 1.142 (AxSYM vs. IMx) and 2.632/0.805 (IMx vs. HPLC). Corresponding mean differences (AxSYM-Comparison Assay) recorded over a 5-50 micromol/l measured range were 1.80, -0.73 and 2.53 micromol/l. AxSYM tHcy FPIA's first rate precision, supported by the complete automation of the AxSYM analyzer, makes it fit for routine use and suitable for laboratories requiring homocysteine high-throughput testing capabilities.

Analytical performance and method comparison study of the total homocysteine fluorescence polarization immunoassay (FPIA) on the AxSYM analyzer.

Accinni R;Campolo J;
2004

Abstract

A fluorescence polarization immunoassay (FPIA) has been commercially released for routine large-scale testing of total homocysteine (tHcy) on the AxSYM analyzer. We evaluated the analytical performance of the AxSYM tHcy FPIA and compared it with the well established high-performance liquid chromatography (HPLC) and IMx tHcy FPIA methods. Homocysteine concentrations were measured by AxSYM and IMx tHcy FPIA and by a rapid isocratic HPLC method with fluorescence detection. Coefficient of variation (CV) of total imprecision for AxSYM tHcy was < or = 5%, mean dilution recovery 102%, analytical sensitivity 0.70 micromol/l and linearity was good up to 1:8 dilution. Spearman rank correlations, rho, were 0.83 (p < 0.0001) for AxSYM vs. HPLC, 0.97 (p < 0.0001) for AxSYM vs. IMx and 0.83 (p < 0.0001) for IMx vs. HPLC. Passing and Bablok regression Y-intercepts and slopes were: 2.944/0.937 (AxSYM vs. HPLC), -0.367/ 1.142 (AxSYM vs. IMx) and 2.632/0.805 (IMx vs. HPLC). Corresponding mean differences (AxSYM-Comparison Assay) recorded over a 5-50 micromol/l measured range were 1.80, -0.73 and 2.53 micromol/l. AxSYM tHcy FPIA's first rate precision, supported by the complete automation of the AxSYM analyzer, makes it fit for routine use and suitable for laboratories requiring homocysteine high-throughput testing capabilities.
2004
homocysteine
analytical methods
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/223393
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact