The work reports on the characterization of high purity epitaxial 4H-SiC grown by liquid phase epitaxy and the performances of Au-Ni/4H-SiC detectors fabricated on the same material. X-ray diffraction and topography as well as I-V, C-V and DLTS measurements are used for the evaluation of the material properties and device characteristics. The X-ray detection abilities are evaluated by pulse-height spectra measurements of the 241Am. Preliminary results of the detector hardness to fast neutron and gamma ray radiations are also reported.
Surface barrier 4H-SiC soft X-ray detecor for hot plasmas diagnostic
E Gombia;C Ferrari;
2012
Abstract
The work reports on the characterization of high purity epitaxial 4H-SiC grown by liquid phase epitaxy and the performances of Au-Ni/4H-SiC detectors fabricated on the same material. X-ray diffraction and topography as well as I-V, C-V and DLTS measurements are used for the evaluation of the material properties and device characteristics. The X-ray detection abilities are evaluated by pulse-height spectra measurements of the 241Am. Preliminary results of the detector hardness to fast neutron and gamma ray radiations are also reported.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.