In two institutions at Rome and Florence we evaluated the clinical sensitivity of two Coulter STKR systems using the NCCLS standard H20-T for leucocyte differential count in a patient population with high prevalence of haematologic abnormalities. Reference ranges of normal leucocytes were obtained on 278 adult subjects. On a population of 455 patient specimens, 200 specimens (44%) were flagged by the STKR because of a distributional abnormality, and 122 (27%) because of a morphological abnormality. Percentage of subtotal agreements between the STKR and the reference manual differential count was 85.4%, with 67.5% full and 20.9% partial agreements. Eight specimens that showed a morphological abnormality with the reference manual differential count were classified as normal by the STKR, with a false normal rate of 6.6%. Analysis of the STKR performance for morphological abnormalities showed acceptable sensitivity (82.0%) and rather low specificity (71.5%), low predictive value of positive results (51.3), high predictive value of negative results (91.5%) and efficiency of 74.3%. The main problems of the STKR differential count were a high rate of false monocyte count, and the misidentification of eosinophilias and low-concentration abnormal cells.

Analysis of leucocyte populations with the Coulter S-Plus STKR as a screening tool for haematological abnormalities

Quarantelli M;
1991

Abstract

In two institutions at Rome and Florence we evaluated the clinical sensitivity of two Coulter STKR systems using the NCCLS standard H20-T for leucocyte differential count in a patient population with high prevalence of haematologic abnormalities. Reference ranges of normal leucocytes were obtained on 278 adult subjects. On a population of 455 patient specimens, 200 specimens (44%) were flagged by the STKR because of a distributional abnormality, and 122 (27%) because of a morphological abnormality. Percentage of subtotal agreements between the STKR and the reference manual differential count was 85.4%, with 67.5% full and 20.9% partial agreements. Eight specimens that showed a morphological abnormality with the reference manual differential count were classified as normal by the STKR, with a false normal rate of 6.6%. Analysis of the STKR performance for morphological abnormalities showed acceptable sensitivity (82.0%) and rather low specificity (71.5%), low predictive value of positive results (51.3), high predictive value of negative results (91.5%) and efficiency of 74.3%. The main problems of the STKR differential count were a high rate of false monocyte count, and the misidentification of eosinophilias and low-concentration abnormal cells.
1991
Inglese
13
1
51
66
Evaluation Studies as Topic Hematologic Diseases/blood/*diagnosis/epidemiology Humans Leukocyte Count/*instrumentation Mass Screening/*methods Predictive Value of Tests Prevalence Reference Values Sensitivity and Specificity
d'Onofrio, G Salvati, A M Berti, P Cappabianca, M P Marsili, G Quarantelli, M Zini, G Casini, C De Philippis, C Mango, G ENGLAND Clin Lab Haematol. 1991;13(1):51-66.
13
info:eu-repo/semantics/article
262
D'Onofrio, G; Salvati, ; A, M; Berti, P; Cappabianca, ; M, P; Marsili, G; Quarantelli, M; Zini, G; Casini, C; De Philippis, C; Mango, G; Et, Al...espandi
01 Contributo su Rivista::01.01 Articolo in rivista
none
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/239307
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact