This paper shows the results of a survey, carried-out in the NMR unit, where a 1.5 T magnet is used, for two years (2006-2008) . The survey was finalized to set-up a suitable methodology to evaluate the safety work conditions of the NMR operating personnel, through the measurement of the magnetic flux density. After, the measured values have been compared with the established limit values.

Methodology for the evaluation of the workers' exposure to magnetic field strength in the NMR unit

Russo A A;
2008

Abstract

This paper shows the results of a survey, carried-out in the NMR unit, where a 1.5 T magnet is used, for two years (2006-2008) . The survey was finalized to set-up a suitable methodology to evaluate the safety work conditions of the NMR operating personnel, through the measurement of the magnetic flux density. After, the measured values have been compared with the established limit values.
2008
INGEGNERIA BIOMEDICA
MRI dosimetry
exposure time
standard and regulation
File in questo prodotto: