This paper proposes the use of a commercial data acquisition board (DAQ) connected to a common personal computer for high accuracy power measurements. In the paper the set up of the PC-based wattmeter is described and an experimental evaluation of its metrological performances is made by means of the comparison with a simultaneous coherent digital sampling wattmeter realized with two high-precision DMMs Hewlett Packard Model 3458 A. In both cases the power source is a Fluke 6100A/E electrical power standard. © 2010 IEEE.
A PC-based wattmeter for high accuracy power measurements
D Di Cara;
2010
Abstract
This paper proposes the use of a commercial data acquisition board (DAQ) connected to a common personal computer for high accuracy power measurements. In the paper the set up of the PC-based wattmeter is described and an experimental evaluation of its metrological performances is made by means of the comparison with a simultaneous coherent digital sampling wattmeter realized with two high-precision DMMs Hewlett Packard Model 3458 A. In both cases the power source is a Fluke 6100A/E electrical power standard. © 2010 IEEE.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.