We developed a compact dual-wavelength dual-channel system for time-resolved diffuse NIR spectroscopy that uses a novel approach based on space-multiplexing (instead of time-multiplexing) of wavelengths, to increase the signal-tonoise ratio and avoid cross-talk. © 2009 SPIE-OSA.
A compact time-resolved system for NIR spectroscopy
Spinelli Lorenzo;Cubeddu Rinaldo;
2009
Abstract
We developed a compact dual-wavelength dual-channel system for time-resolved diffuse NIR spectroscopy that uses a novel approach based on space-multiplexing (instead of time-multiplexing) of wavelengths, to increase the signal-tonoise ratio and avoid cross-talk. © 2009 SPIE-OSA.File in questo prodotto: