Leakage current and capacity dependence on temperature have been measured in the range - 30°C < t < 20°C for several silicon strip detectors. The experimental setup is composed of a copper test box enclosing 16 current-to-voltage converters and a refrigerator to set and control the required temperatures. These measurements were performed for noise minimization for the space calorimeter of the WiZard experiment to be operated on the American space station Freedom. The results show that the overall noise of the electronic readout chain may be significantly reduced by operating the calorimeter at low temperature. © 1991.

Leakage current and capacity variation with temperature in silicon detectors of a space calorimeter

Brunetti MT;
1991

Abstract

Leakage current and capacity dependence on temperature have been measured in the range - 30°C < t < 20°C for several silicon strip detectors. The experimental setup is composed of a copper test box enclosing 16 current-to-voltage converters and a refrigerator to set and control the required temperatures. These measurements were performed for noise minimization for the space calorimeter of the WiZard experiment to be operated on the American space station Freedom. The results show that the overall noise of the electronic readout chain may be significantly reduced by operating the calorimeter at low temperature. © 1991.
File in questo prodotto:
File Dimensione Formato  
prod_283902-doc_93223.pdf

non disponibili

Descrizione: Leakage current and capacity variation with temperature in silicon detectors of a space calorimeter
Dimensione 299.63 kB
Formato Adobe PDF
299.63 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/253301
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
  • ???jsp.display-item.citation.isi??? ND
social impact