Two different approaches were tested in this study to suppress sulfide interference in the silicomolybdate method of silica determination, namely: oxidation of sulfide to elemental sulfur; and conversion of the sulfide species to H2S and stripping by air bubbling. Based on the results obtained the latter approach is deemed to represent the better and quicker option to eliminate sulfide interference in silica colorimetric analysis.

A procedure for eliminating sulfide interference on silica colorimetric analysis

Lelli M;
2014

Abstract

Two different approaches were tested in this study to suppress sulfide interference in the silicomolybdate method of silica determination, namely: oxidation of sulfide to elemental sulfur; and conversion of the sulfide species to H2S and stripping by air bubbling. Based on the results obtained the latter approach is deemed to represent the better and quicker option to eliminate sulfide interference in silica colorimetric analysis.
2014
Istituto di Geoscienze e Georisorse - IGG - Sede Pisa
sulfide interference
colorimetric analysis
silica
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/255722
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact