Background. Preclinical perfusion studies are useful for the improvement of diagnosis and therapy in dermatologic, cardiovascular and rheumatic human diseases. The Laser Doppler Perfusion Imaging (LDPI) technique has been used to evaluate superficial alterations of the skin microcirculation in surgically induced murine hindlimb ischemia. We assessed the reproducibility and the accuracy of LDPI acquisitions and identified several critical factors that could affect LDPI measurements in mice. Methods. Twenty mice were analysed. Statistical standardisation and a repeatability and reproducibility analysis were performed on mouse perfusion signals with respect to differences in body temperature, the presence or absence of hair, the type of anaesthesia used for LDPI measurements and the position of the mouse body. Results. We found excellent correlations among measurements made by the same operator (i.e., repeatability) under the same experimental conditions and by two different operators (i.e., reproducibility). A Bland-Altman analysis showed the absence of bias in repeatability (p = 0.29) or reproducibility (p = 0.89). The limits of agreement for repeatability were -0.357 and -0.033, and for reproducibility, they were -0.270 and 0.238. Significant differences in perfusion values were observed in different experimental groups. Conclusions. Different experimental conditions must be considered as a starting point for the evaluation of new drugs and strategic therapies.

Repeatability, Reproducibility and Standardisation of a Laser Doppler Imaging Technique for the Evaluation of Normal Mouse Hindlimb Perfusion

Raffaele Liuzzi;Sara Gargiulo;Matteo Gramanzini;Marcello Mancini;
2013

Abstract

Background. Preclinical perfusion studies are useful for the improvement of diagnosis and therapy in dermatologic, cardiovascular and rheumatic human diseases. The Laser Doppler Perfusion Imaging (LDPI) technique has been used to evaluate superficial alterations of the skin microcirculation in surgically induced murine hindlimb ischemia. We assessed the reproducibility and the accuracy of LDPI acquisitions and identified several critical factors that could affect LDPI measurements in mice. Methods. Twenty mice were analysed. Statistical standardisation and a repeatability and reproducibility analysis were performed on mouse perfusion signals with respect to differences in body temperature, the presence or absence of hair, the type of anaesthesia used for LDPI measurements and the position of the mouse body. Results. We found excellent correlations among measurements made by the same operator (i.e., repeatability) under the same experimental conditions and by two different operators (i.e., reproducibility). A Bland-Altman analysis showed the absence of bias in repeatability (p = 0.29) or reproducibility (p = 0.89). The limits of agreement for repeatability were -0.357 and -0.033, and for reproducibility, they were -0.270 and 0.238. Significant differences in perfusion values were observed in different experimental groups. Conclusions. Different experimental conditions must be considered as a starting point for the evaluation of new drugs and strategic therapies.
2013
Istituto di Biostrutture e Bioimmagini - IBB - Sede Napoli
Laser Doppler Perfusion Imaging
hindlimb perfusion
mice; microcirculation
standardisation
File in questo prodotto:
File Dimensione Formato  
prod_275306-doc_89347.pdf

solo utenti autorizzati

Descrizione: Repeatability, Reproducibility and Standardisation of a Laser Doppler Imaging .... Mouse Hindlimb Perfusion
Dimensione 564.78 kB
Formato Adobe PDF
564.78 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/258078
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 16
  • ???jsp.display-item.citation.isi??? 15
social impact