Various clinical procedures, such as cardiovascular surgery or extracorporeal blood purification, involve systemic anticoagulation using heparin. High concentrations of circulating heparin require neutralization due to possible serious bleeding complications. The intravenous administration of the heparin antagonist protamine sulfate is routinely clinically performed, but is frequently associated with adverse reactions. Therefore, there is a need for a valid and safe alternative to achieve extracorporeal heparin removal from blood or plasma, such as a filter, a matrix, or an adsorbent. Here, we describe the development of a macroporous poly(2-hydroxyethyl methacrylate)-based monolithic cryogel functionalized with l-lysine (pHEMA-lys) and the characterization of its selective heparin adsorption. The maximum binding capacity was quantified in vitro using aqueous and serum solutions under static and dynamic conditions, and fresh human plasma under static conditions. The pHEMA-lys bound 40,500 IU and 32,500 IU heparin/g cryogel at the equilibrium in aqueous solution and 50% serum, respectively. In human plasma spiked with 100 IU/mL of heparin, the binding was still highly efficient (4330 IU/g cryogel after 30 min, i.e., 87% of the initial concentration). The cryogels showed good blood compatibility, as indicated by negligible adsorption of albumin, antithrombin III, and total protein, and may thus be suitable for extracorporeal heparin removal. © 2014 Wiley Periodicals, Inc.

Chemically modified poly(2-hydroxyethyl methacrylate) cryogel for the adsorption of heparin

Mecca T;Cunsolo F;
2014

Abstract

Various clinical procedures, such as cardiovascular surgery or extracorporeal blood purification, involve systemic anticoagulation using heparin. High concentrations of circulating heparin require neutralization due to possible serious bleeding complications. The intravenous administration of the heparin antagonist protamine sulfate is routinely clinically performed, but is frequently associated with adverse reactions. Therefore, there is a need for a valid and safe alternative to achieve extracorporeal heparin removal from blood or plasma, such as a filter, a matrix, or an adsorbent. Here, we describe the development of a macroporous poly(2-hydroxyethyl methacrylate)-based monolithic cryogel functionalized with l-lysine (pHEMA-lys) and the characterization of its selective heparin adsorption. The maximum binding capacity was quantified in vitro using aqueous and serum solutions under static and dynamic conditions, and fresh human plasma under static conditions. The pHEMA-lys bound 40,500 IU and 32,500 IU heparin/g cryogel at the equilibrium in aqueous solution and 50% serum, respectively. In human plasma spiked with 100 IU/mL of heparin, the binding was still highly efficient (4330 IU/g cryogel after 30 min, i.e., 87% of the initial concentration). The cryogels showed good blood compatibility, as indicated by negligible adsorption of albumin, antithrombin III, and total protein, and may thus be suitable for extracorporeal heparin removal. © 2014 Wiley Periodicals, Inc.
2014
Istituto di Chimica Biomolecolare - ICB - Sede Pozzuoli
adsorption
hemocompatibility
heparin
pHEMA
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/258315
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 11
  • ???jsp.display-item.citation.isi??? ND
social impact