In this work we investigate, for the first time, the performances of a system based on hydrogenated amorphous silicon photosensors for the detection of Ochratoxin A. The sensor is a n-type/intrinsic/p-type amorphous silicon stacked structure deposited on a glass substrate. The mycotoxin is deposited on a thin layer chromatographic plate and aligned with the sensor. An ultraviolet radiation excites the ochratoxin A, whose fluorescence produces a photocurrent in the sensor. The photocurrent value is proportional to the deposited mycotoxin quantity. An excellent linearity of the detector response over more than two orders of magnitude of ochratoxin A amount is observed. The minimum detected mycotoxin quantity is equal to 0.1 ng, suggesting that the presented detection system could be a good candidate to perform rapid and analytical ochratoxin A analysis in different kind of samples.

An alternative system for mycotoxin detection based on amorphous silicon sensors - art. no. 65891F

Ricelli A;
2007

Abstract

In this work we investigate, for the first time, the performances of a system based on hydrogenated amorphous silicon photosensors for the detection of Ochratoxin A. The sensor is a n-type/intrinsic/p-type amorphous silicon stacked structure deposited on a glass substrate. The mycotoxin is deposited on a thin layer chromatographic plate and aligned with the sensor. An ultraviolet radiation excites the ochratoxin A, whose fluorescence produces a photocurrent in the sensor. The photocurrent value is proportional to the deposited mycotoxin quantity. An excellent linearity of the detector response over more than two orders of magnitude of ochratoxin A amount is observed. The minimum detected mycotoxin quantity is equal to 0.1 ng, suggesting that the presented detection system could be a good candidate to perform rapid and analytical ochratoxin A analysis in different kind of samples.
2007
Inglese
Smart Sensors, Actuators, and MEMS III Book Series: PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS (SPIE)
Conference on Smart Sensors, Actuators and MEMS III
6589
F5891
F5891
6
978-0-8194-6717-1
Sì, ma tipo non specificato
MAY 02-04, 2007
Maspalomas, SPAIN
amorphous silicon
toxin detection
ochratoxin A
photosensor
1
none
Caputo, D.; de Cesare, G.; de Rossi, P.; Fanelli, C.; Nascetti, A.; Ricelli, A.; Scipinotti, R.
273
info:eu-repo/semantics/conferenceObject
04 Contributo in convegno::04.01 Contributo in Atti di convegno
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/275211
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? ND
social impact