This model study evaluates the effect of pump characteristics and cardiovascular data on hemodynamics in atrio-aortic VAD assistance. The model includes a computational circulatory sub-model and an electrical sub-model representing two rotary blood pumps through their pressure-flow characteristics. The first is close to a pressure generator-PG (average flow sensitivity to pressure variations, -0.047 l mmHg(-1)); the second is closer to a flow generator-FG (average flow sensitivity to pressure variations, -0.0097 l mmHg(-1)). Interaction with VAD was achieved by means of two interfaces, behaving as impedance transformers. The model was verified by use of literature data and VAD onset conditions were used as a control for the experiments. Tests compared the two pumps, at constant pump speed, in different ventricular and circulatory conditions: maximum ventricular elastance (0.44-0.9 mmHg cm(-3)), systemic peripheral resistance (781-1200 g cm(-4) s(-1)), ventricular diastolic compliance C (p) (5-10-50 cm(3) mmHg(-1)), systemic arterial compliance (0.9-1.8 cm(3) mmHg(-1)). Analyzed variables were: arterial and venous pressures, flows, ventricular volume, external work, and surplus hemodynamic energy (SHE). The PG pump generated the highest SHE under almost all conditions, in particular for higher C (p) (+50 %). PG pump flow is also the most sensitive to E (max) and C (p) changes (-26 and -33 %, respectively). The FG pump generally guarantees higher external work reduction (54 %) and flow less dependent on circulatory and ventricular conditions. The results are evidence of the importance of pump speed regulation with changing ventricular conditions. The computational sub-model will be part of a hydro-numerical model, including autonomic controls, designed to test different VADs.

Continuous-flow pump model study: the effect on pump performance of pump characteristics and cardiovascular conditions

Ferrari Gianfranco;Fresiello Libera;Di Molfetta Arianna;
2013

Abstract

This model study evaluates the effect of pump characteristics and cardiovascular data on hemodynamics in atrio-aortic VAD assistance. The model includes a computational circulatory sub-model and an electrical sub-model representing two rotary blood pumps through their pressure-flow characteristics. The first is close to a pressure generator-PG (average flow sensitivity to pressure variations, -0.047 l mmHg(-1)); the second is closer to a flow generator-FG (average flow sensitivity to pressure variations, -0.0097 l mmHg(-1)). Interaction with VAD was achieved by means of two interfaces, behaving as impedance transformers. The model was verified by use of literature data and VAD onset conditions were used as a control for the experiments. Tests compared the two pumps, at constant pump speed, in different ventricular and circulatory conditions: maximum ventricular elastance (0.44-0.9 mmHg cm(-3)), systemic peripheral resistance (781-1200 g cm(-4) s(-1)), ventricular diastolic compliance C (p) (5-10-50 cm(3) mmHg(-1)), systemic arterial compliance (0.9-1.8 cm(3) mmHg(-1)). Analyzed variables were: arterial and venous pressures, flows, ventricular volume, external work, and surplus hemodynamic energy (SHE). The PG pump generated the highest SHE under almost all conditions, in particular for higher C (p) (+50 %). PG pump flow is also the most sensitive to E (max) and C (p) changes (-26 and -33 %, respectively). The FG pump generally guarantees higher external work reduction (54 %) and flow less dependent on circulatory and ventricular conditions. The results are evidence of the importance of pump speed regulation with changing ventricular conditions. The computational sub-model will be part of a hydro-numerical model, including autonomic controls, designed to test different VADs.
2013
Lumped parameter model
Computational model
Physical model
Heart assist device
Rotary blood pump
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/278188
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 5
  • ???jsp.display-item.citation.isi??? ND
social impact