Scanning Mobility Particle Sizer (SMPS) spectrometers are subject to several restrictions when radioactive sources are employed to bring the aerosol to a steady state charge distribution. An alternative solution is represented by neutralizers based on soft X-rays. The present research investigates, for the first time, the outcome of a combination of the soft X-ray advanced aerosol neutralizer (TSI model 3087) with a Grimm SMPS (L-DMA, mod. 5400).

Application of the soft X-ray TSI Advanced Aerosol Neutralizer to aerosol measurements made by a Grimm SMPS

Nicosia A;Belosi F;Santachiara G;Prodi F;
2015

Abstract

Scanning Mobility Particle Sizer (SMPS) spectrometers are subject to several restrictions when radioactive sources are employed to bring the aerosol to a steady state charge distribution. An alternative solution is represented by neutralizers based on soft X-rays. The present research investigates, for the first time, the outcome of a combination of the soft X-ray advanced aerosol neutralizer (TSI model 3087) with a Grimm SMPS (L-DMA, mod. 5400).
2015
Istituto di Scienze dell'Atmosfera e del Clima - ISAC
DMA
nanoparticle
File in questo prodotto: