Monitoring the water content and detecting the presence of soluble salts in concrete is a key issue for its maintenance. Evanescent field dielectrometry, originally developed for the diagnostics of frescoes and mural paintings, is proposed as a tool for monitoring the decay of cement-based materials. A measuring system, based on a scalar network analyzer and a resonant probe, has been realized and tested on concrete samples taken from historical buildings in France or purposely developed in the laboratory. Measurements on water-saturated and oven-dry samples provide the basis for calibrating the instrument for on site monitoring of concrete historical buildings, sculptures and cement-based artifacts.
Moisture and salt monitoring in concrete by evanescent field dielectrometry
Riminesi C;Olmi R
2017
Abstract
Monitoring the water content and detecting the presence of soluble salts in concrete is a key issue for its maintenance. Evanescent field dielectrometry, originally developed for the diagnostics of frescoes and mural paintings, is proposed as a tool for monitoring the decay of cement-based materials. A measuring system, based on a scalar network analyzer and a resonant probe, has been realized and tested on concrete samples taken from historical buildings in France or purposely developed in the laboratory. Measurements on water-saturated and oven-dry samples provide the basis for calibrating the instrument for on site monitoring of concrete historical buildings, sculptures and cement-based artifacts.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.