Objectives We developed a wearable sensor for the real time measurement of sweat rate in localized areas of the human body. This sensor represents the first step in the development of a wearable sensor network capable to estimate the global sweat rate via an ad hoc algorithm. Such device would be used to monitor athletes' hydration status during training and improve their performances. Equipment and methods For this study, we tested our sensor on thirteen football players during a cycling test on a cycle ergometer. The sweat rate sensor was compared to a medical device that, although measuring a different physiological process, provides discrete data based on the same working principle, i.e. the diffusion of the water vapour emitted from the skin. Results Our sensor has a working range up to 400 g/m2·h. The statistical analysis and the Bland-Altman plot proved that our sensor is comparable to the medical device used as gold standard. At low sweat rate, the bias is 3.4 g/m2·h with a standard deviation of 7.6 g/m2·h. At maximum sweat rates, the bias is 2.3 g/m2·h with a standard deviation 6.9 g/m2·h. The P-values for the Bland-Altman plots at low and maximum sweat rate (0.1331 and 0.2477 obtained by Kolmogorov-Smirnov test, respectively) allow the hypothesis that there is a significant difference between our sweat rate sensor and the medical device to be rejected. Conclusion We presented a prototype of a wearable sweat rate sensor for localized measurements. The trials on thirteen athletes proved that the performance of our sensor is comparable to that of a commercial medical device. This sweat rate sensor can provide valuable information on athletes' hydration status.

A wearable sweat rate sensor to monitor the athletes' performance during training

Salvo P;Pingitore A;Barbini A;
2018

Abstract

Objectives We developed a wearable sensor for the real time measurement of sweat rate in localized areas of the human body. This sensor represents the first step in the development of a wearable sensor network capable to estimate the global sweat rate via an ad hoc algorithm. Such device would be used to monitor athletes' hydration status during training and improve their performances. Equipment and methods For this study, we tested our sensor on thirteen football players during a cycling test on a cycle ergometer. The sweat rate sensor was compared to a medical device that, although measuring a different physiological process, provides discrete data based on the same working principle, i.e. the diffusion of the water vapour emitted from the skin. Results Our sensor has a working range up to 400 g/m2·h. The statistical analysis and the Bland-Altman plot proved that our sensor is comparable to the medical device used as gold standard. At low sweat rate, the bias is 3.4 g/m2·h with a standard deviation of 7.6 g/m2·h. At maximum sweat rates, the bias is 2.3 g/m2·h with a standard deviation 6.9 g/m2·h. The P-values for the Bland-Altman plots at low and maximum sweat rate (0.1331 and 0.2477 obtained by Kolmogorov-Smirnov test, respectively) allow the hypothesis that there is a significant difference between our sweat rate sensor and the medical device to be rejected. Conclusion We presented a prototype of a wearable sweat rate sensor for localized measurements. The trials on thirteen athletes proved that the performance of our sensor is comparable to that of a commercial medical device. This sweat rate sensor can provide valuable information on athletes' hydration status.
2018
Istituto di Fisiologia Clinica - IFC
Istituto Nazionale di Ottica - INO
Hydration
Personalized training
Sports performance
Sweat rate
Wearable sensor
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/334870
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 33
  • ???jsp.display-item.citation.isi??? ND
social impact