The aim was to assess the influence of a biventricular assist device (BVAD) on ventricular energetics parameters (external work, oxygen consumption, cardiac mechanical efficiency) for both ventricles, when mechanical ventilation was applied. The experiments were performed using a computer simulator of cardiovascular system (CARDIOSIM) after modelling a pathological state of the left ventricle (E(v)Left = 0. 9 mmHg cm(-3) and increasing pulmonary resistance (Rap = 0.3 mmHg cm(-3 s). The effect of mechanical ventilation was mean intrathoracic pressure changes from 0 to +5 mmHg. This simulation showed that application of BVAD for both ventricles reduces external work and that this effect is stressed by positive intrathoracic pressure, reduces cardiac mechanical efficiency that is quite insensitive to intrathoracic pressure and increases oxygen consumption, which is reduced by positive intrathoracic pressure. The increase of potential energy at the onset of BVAD evidences a rightwards shift of ventricular work cycle (unloading of the ventricles). In general, positive intrathoracic pressure during BVAD assistance adversely affects ventricular energetics.

Energetic parameter changes with mechanical ventilation in conjunction with BVAD assistance

De Lazzari C;Ferrari G;Clemente F;
2002

Abstract

The aim was to assess the influence of a biventricular assist device (BVAD) on ventricular energetics parameters (external work, oxygen consumption, cardiac mechanical efficiency) for both ventricles, when mechanical ventilation was applied. The experiments were performed using a computer simulator of cardiovascular system (CARDIOSIM) after modelling a pathological state of the left ventricle (E(v)Left = 0. 9 mmHg cm(-3) and increasing pulmonary resistance (Rap = 0.3 mmHg cm(-3 s). The effect of mechanical ventilation was mean intrathoracic pressure changes from 0 to +5 mmHg. This simulation showed that application of BVAD for both ventricles reduces external work and that this effect is stressed by positive intrathoracic pressure, reduces cardiac mechanical efficiency that is quite insensitive to intrathoracic pressure and increases oxygen consumption, which is reduced by positive intrathoracic pressure. The increase of potential energy at the onset of BVAD evidences a rightwards shift of ventricular work cycle (unloading of the ventricles). In general, positive intrathoracic pressure during BVAD assistance adversely affects ventricular energetics.
2002
Circulatory model
Lumped parameters
Variable elastance model
ventricular energetics
intrathoracic pressure
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/336526
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 4
  • ???jsp.display-item.citation.isi??? ND
social impact