The analytical performances of a lateral flow immunoassay (AFLA-V AQUA(TM), Vicam a Waters Business) for the determination of aflatoxins in maize were evaluated according to Commission Regulation (EU) No. 519/ 2014. This case study showed how a critical evaluation of validation results can provide a quite informative picture of kit performances for its use for different purposes. The validation study, performed at aflatoxin B1 levels encompassing EU maximum permitted levels (up to 8 ug kg-1), provided information on the method precision profile, cut off values, and false suspect and false negative rates. The total precision expressed as relative standard deviation varied from 14 to 29% for contaminated samples. Cut off values calculated considering 2 ug kg-1 or 4 mg kg-1 as the screening target concentration were 1.24 and 2.18 ug kg-1, respectively, whereas the resultant false suspect rate for blanks was 42% and 8%, respectively. The false negative rate for samples containing the analyte at higher concentrations was found to be lower than 1%. Analysis of naturally contaminated maize samples performed by the lateral flow immunoassay and the AOAC official method revealed a good correlation (parameters of the linear regression: r = 0.97, slope = 0.96) between the obtained results. A satisfactory agreement between reference values and test results was also obtained in the analysis of reference materials. Finally, a QC protocol based on guidelines suggested in the EU regulation was designed to verify kit performances over time and in new production lots.

Validation of a lateral flow immunoassay for the rapid determination of aflatoxins in maize by solvent free extraction

Biancamaria Ciasca;Michelangelo Pascale;
2017

Abstract

The analytical performances of a lateral flow immunoassay (AFLA-V AQUA(TM), Vicam a Waters Business) for the determination of aflatoxins in maize were evaluated according to Commission Regulation (EU) No. 519/ 2014. This case study showed how a critical evaluation of validation results can provide a quite informative picture of kit performances for its use for different purposes. The validation study, performed at aflatoxin B1 levels encompassing EU maximum permitted levels (up to 8 ug kg-1), provided information on the method precision profile, cut off values, and false suspect and false negative rates. The total precision expressed as relative standard deviation varied from 14 to 29% for contaminated samples. Cut off values calculated considering 2 ug kg-1 or 4 mg kg-1 as the screening target concentration were 1.24 and 2.18 ug kg-1, respectively, whereas the resultant false suspect rate for blanks was 42% and 8%, respectively. The false negative rate for samples containing the analyte at higher concentrations was found to be lower than 1%. Analysis of naturally contaminated maize samples performed by the lateral flow immunoassay and the AOAC official method revealed a good correlation (parameters of the linear regression: r = 0.97, slope = 0.96) between the obtained results. A satisfactory agreement between reference values and test results was also obtained in the analysis of reference materials. Finally, a QC protocol based on guidelines suggested in the EU regulation was designed to verify kit performances over time and in new production lots.
2017
Istituto di Scienze delle Produzioni Alimentari - ISPA
: aflatoxins
screening methods
validation
lateral flow devices
maize
mycotoxins
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/336684
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact