The efficient use of computed tomography (CT) and magnetic resonance imaging (MRI) equipment necessitates establishing adequate quality-control (QC) procedures. In particular, the accuracy of slice thickness (ST) requires scan exploration of phantoms containing test objects (plane, cone or spiral). To simplify such procedures, a novel phantom and a computerised LabView-based procedure have been devised, enabling determination of full width at half maximum (FWHM) in real time.

Slice-thickness evaluation in CT and MRI: an alternative computerised procedure

Causa F;
2012

Abstract

The efficient use of computed tomography (CT) and magnetic resonance imaging (MRI) equipment necessitates establishing adequate quality-control (QC) procedures. In particular, the accuracy of slice thickness (ST) requires scan exploration of phantoms containing test objects (plane, cone or spiral). To simplify such procedures, a novel phantom and a computerised LabView-based procedure have been devised, enabling determination of full width at half maximum (FWHM) in real time.
2012
Istituto di fisica del plasma - IFP - Sede Milano
Quality control
Computed tomography
Magnetic resonance
Slice thickness
File in questo prodotto:
File Dimensione Formato  
prod_378816-doc_128189.pdf

solo utenti autorizzati

Descrizione: Articolo pubblicato
Tipologia: Versione Editoriale (PDF)
Dimensione 634.79 kB
Formato Adobe PDF
634.79 kB Adobe PDF   Visualizza/Apri   Richiedi una copia