Computer simulation of a cardiovascular system enabled us to predict the effects of simultaneous application of mechanical ventilation (MV) and intraaortic ballon pumping (IABP) on ventricular energetics. External work (EW), pressure-volume area (PVA), potential energy (PE) and cardiac mechanical efficiency (CME) were calculated. Nummerical simulation showed that changes of positive intrathoracic pressure have a considerable effect on left and right ventricular EW, PE, PVA and CME, whether IABP is used or not. The right ventricular energetics was much less influenced by systemic resistance (Ras) changes than the left ventricular one. Simultaneous application of IABP and MV showed a remarkable effect on left ventricular EW. The net result was reversed sensitivity to pulmonary resistance (Rap) and reduced sensitivity to Ras. PVA was generally reduced, while CME is increased by simultaneous presence of IABP and MV. The sensitivity of CME to Rap and Ras variation was diminished in this situation.

Ventricular energetics during mechanical ventilation and intraaortic balloon pumping - Computer simulation

De Lazzari C;Ferrari G;Clemente F;
2001

Abstract

Computer simulation of a cardiovascular system enabled us to predict the effects of simultaneous application of mechanical ventilation (MV) and intraaortic ballon pumping (IABP) on ventricular energetics. External work (EW), pressure-volume area (PVA), potential energy (PE) and cardiac mechanical efficiency (CME) were calculated. Nummerical simulation showed that changes of positive intrathoracic pressure have a considerable effect on left and right ventricular EW, PE, PVA and CME, whether IABP is used or not. The right ventricular energetics was much less influenced by systemic resistance (Ras) changes than the left ventricular one. Simultaneous application of IABP and MV showed a remarkable effect on left ventricular EW. The net result was reversed sensitivity to pulmonary resistance (Rap) and reduced sensitivity to Ras. PVA was generally reduced, while CME is increased by simultaneous presence of IABP and MV. The sensitivity of CME to Rap and Ras variation was diminished in this situation.
2001
LVAD
BVAD
Emax
Ventricular energetics
Simulation
Variable elastance model
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/338674
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 18
  • ???jsp.display-item.citation.isi??? ND
social impact