This study is aimed at optimizing the CO2 removal in the ventilator prototype for neonatal Total Liquid Ventilation (TLV) designed by our research group. A computational approach was adopted to evaluate carbon dioxide exchange between the liquid perfluorocarbon and the oxygenating gas mixture, as a function of the geometrical parameters of the device. The computational model proved to be a good tool to simulate gas transfer and allowed the design of a new ventilator prototype, currently undergoing in vivo trials.

Computational model of CO2 removal in a prototype for neonatal Total Liquid Ventilation

Arlati S;
2014

Abstract

This study is aimed at optimizing the CO2 removal in the ventilator prototype for neonatal Total Liquid Ventilation (TLV) designed by our research group. A computational approach was adopted to evaluate carbon dioxide exchange between the liquid perfluorocarbon and the oxygenating gas mixture, as a function of the geometrical parameters of the device. The computational model proved to be a good tool to simulate gas transfer and allowed the design of a new ventilator prototype, currently undergoing in vivo trials.
2014
9788855532754
Total Liquid Ventilation
Ventilator Prototype
gas exchange
computational model
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/344108
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact