The Short-Time Retractable Instrumented Kalorimeter Experiment (STRIKE) diagnostic will be used to characterize the SPIDER negative ion beam during short pulse operation (< 10 s) to verify the degree of attainment of ITER requirements about the maximum allowed beam non-uniformity ( 10% is required). After a preliminary design developed in the last few years, the complete STRIKE diagnostic system has been recently subjected to a final review and is now ready for construction. The main components of the system are: an array of 16 tiles made of unidirectional Carbon Fibre Composite (CFC) with the fibre direction parallel to the thickness, which will be exposed to the high power density beam (20 MW/m(2) peak) exiting from the SPIDER Beam Source; a supporting and positioning system, based on a set of structures made of stainless steel, with controlled moving systems, to be installed within the SPIDER vacuum vessel; a set of thermal, electric and thermo-graphic sensors to thoroughly characterize the beam. The development of the engineering design of the supporting and positioning system and an overview of the complete diagnostic system are given.

Final design of the diagnostic calorimeter for the negative ion source SPIDER

Brombin M;Cervaro V;Dalla Palma M;Peruzzo S;Pasqualotto R;Serianni G
2017

Abstract

The Short-Time Retractable Instrumented Kalorimeter Experiment (STRIKE) diagnostic will be used to characterize the SPIDER negative ion beam during short pulse operation (< 10 s) to verify the degree of attainment of ITER requirements about the maximum allowed beam non-uniformity ( 10% is required). After a preliminary design developed in the last few years, the complete STRIKE diagnostic system has been recently subjected to a final review and is now ready for construction. The main components of the system are: an array of 16 tiles made of unidirectional Carbon Fibre Composite (CFC) with the fibre direction parallel to the thickness, which will be exposed to the high power density beam (20 MW/m(2) peak) exiting from the SPIDER Beam Source; a supporting and positioning system, based on a set of structures made of stainless steel, with controlled moving systems, to be installed within the SPIDER vacuum vessel; a set of thermal, electric and thermo-graphic sensors to thoroughly characterize the beam. The development of the engineering design of the supporting and positioning system and an overview of the complete diagnostic system are given.
2017
Istituto gas ionizzati - IGI - Sede Padova
Inglese
123
768
772
5
https://www.sciencedirect.com/science/article/pii/S0920379617305380
Sì, ma tipo non specificato
NBI
Calorimeter
CFC
Thermal camera
Electronic ISSN: 1873-7196 / The work leading to this publication has been funded partially by Fusion for Energy under the Contract F4E-RFXPMS A-WP-2015. / http://biblioproxy.cnr.it:2084/full_record.do?product=WOS&search_mode=GeneralSearch&qid=1&SID=F18JMQJUVze97M1uBTd&page=1&doc=1
11
info:eu-repo/semantics/article
262
Rizzolo, A; Tollin, M; Brombin, M; Cervaro, V; Dalla Palma, M; De Muri, M; Fasolo, D; Franchin, L; Peruzzo, S; Pimazzoni, A; Pasqualotto, R; Serianni,...espandi
01 Contributo su Rivista::01.01 Articolo in rivista
restricted
   EU Fusion for ITER Applications
   EUFORIA
   FP7
   211804
File in questo prodotto:
File Dimensione Formato  
prod_384615-doc_131587.pdf

solo utenti autorizzati

Descrizione: Final design of the diagnostic calorimeter for the negative ion source SPIDER
Tipologia: Versione Editoriale (PDF)
Dimensione 1.99 MB
Formato Adobe PDF
1.99 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/351141
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 15
social impact