A particular procedure is suggested to implement a high-range-resolution pulse-echo ultrasound technique by means of a frequency-domain exploration of the tested device and a successive Fourier inverse transformation of the measured data. The improvement with respect to conventional methods is estimated, and an experimental setup based on the use of a commercially available vectorial network analyzer is presented. © 1994 IEEE
An improved range-resolution pulse-echo technique by means of vectorial-network-analyzer-based frequency-domain testing
1994
Abstract
A particular procedure is suggested to implement a high-range-resolution pulse-echo ultrasound technique by means of a frequency-domain exploration of the tested device and a successive Fourier inverse transformation of the measured data. The improvement with respect to conventional methods is estimated, and an experimental setup based on the use of a commercially available vectorial network analyzer is presented. © 1994 IEEEFile in questo prodotto:
| File | Dimensione | Formato | |
|---|---|---|---|
|
prod_409613-doc_144036.pdf
solo utenti autorizzati
Descrizione: An improved range-resolution pulse-echo technique by means of vectorial-network-analyzer-based frequency-domain testing