The oxygen concentration in the headspace of a large number of liquid food containers is assessed with non-intrusive diode laser absorption spectroscopy. The results are compared to those given by a traditional intrusive measurement method. The upper portions of the containers are made of translucent but non-transparent plastic materials, through which the laser light is diffusively transmitted. Measurements are performed both on packages with modified atmosphere and with normal air conditions. It is concluded that the novel non-intrusive technique provides oxygen concentrations that generally differ by less than one percentage point from the reference sensor values. (C) 2012 Elsevier Ltd. All rights reserved.
Non-intrusive headspace gas measurements by laser spectroscopy - Performance validation by a reference sensor
Cocola Lorenzo;
2012
Abstract
The oxygen concentration in the headspace of a large number of liquid food containers is assessed with non-intrusive diode laser absorption spectroscopy. The results are compared to those given by a traditional intrusive measurement method. The upper portions of the containers are made of translucent but non-transparent plastic materials, through which the laser light is diffusively transmitted. Measurements are performed both on packages with modified atmosphere and with normal air conditions. It is concluded that the novel non-intrusive technique provides oxygen concentrations that generally differ by less than one percentage point from the reference sensor values. (C) 2012 Elsevier Ltd. All rights reserved.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.