We propose a simple and reliable solid phantom for mimicking realistic localized absorption changes within a diffusive medium. The phantom is based on a solid matrix holding a movable black inclusion embedded in a rod. Translating the rod parallel to the phantom surface, the inhomogeneity can be positioned beneath the source-detector pair (perturbed case) or far from it (unperturbed case). Examples of time-resolved transmittance measurements and time-resolved reflectance scans are shown to demonstrate the properties and the versatility of the phantom.

Design and construction of a solid switchable phantom for diffuse optical imaging

Pifferi Antonio;Spinelli Lorenzo;Farina Andrea;Torricelli Alessandro;
2015

Abstract

We propose a simple and reliable solid phantom for mimicking realistic localized absorption changes within a diffusive medium. The phantom is based on a solid matrix holding a movable black inclusion embedded in a rod. Translating the rod parallel to the phantom surface, the inhomogeneity can be positioned beneath the source-detector pair (perturbed case) or far from it (unperturbed case). Examples of time-resolved transmittance measurements and time-resolved reflectance scans are shown to demonstrate the properties and the versatility of the phantom.
2015
Istituto di fotonica e nanotecnologie - IFN
Phantom
diffuse optics
time-resolved spectroscopy
absorption
scattering
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/377245
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 0
social impact