The remote, timely and in-field detection of sulfured additives in natural gas pipelines is a challenge for environmental, commercial and safety reasons. Moreover, the constant control of the level of odorants in a pipeline is required by law to prevent explosions and accidents. Currently, the detection of the most common odorants (THT = tetrahydrothiophane; TBM = tertiary butyl mercaptan) added to natural gas streams in pipelines is made in situ by using portable gas chromatography apparatuses. In this study, we report the analysis of the ultraviolet spectra obtained by a customized ultraviolet spectrophotometer, named Spectra, for the in-field detection of thiophane and tertiary butyl mercaptan. Spectra were conceived to accomplish the remote analysis of odorants in the pipelines of the natural gas stream through the adoption of technical solutions aimed to adapt a basic bench ultraviolet spectrophotometer to the in-field analysis of gases. The remotely controlled system acquires spectra continuously, performing the quantitative determination of odorants and catching systemic or accidental variations of the gaseous mixture in different sites of the pipeline. The analysis of the experimental spectra was carried out also through theoretical quantum mechanical approaches aimed to detect and to correctly assign the nature of the intrinsic electronic transitions of the two odorants, thiophane and tertiary butyl mercaptan, that cause the ultraviolet absorptions. So far, these theoretical aspects have never been studied before. The absorption maxima of thiophane and tertiary butyl mercaptan spectra were computationally simulated through the usage of selected molecular models with satisfactory results. The good matches between the experimental and theoretical datasets corroborate the reliability of the collected data. During the tests, unexpected pollutants and accidental malfunctions have been detected and also identified by Spectra, making this instrument suitable for many purposes.

Odorant Monitoring in Natural Gas Pipelines Using Ultraviolet-Visible Spectroscopy

Manca G
2021

Abstract

The remote, timely and in-field detection of sulfured additives in natural gas pipelines is a challenge for environmental, commercial and safety reasons. Moreover, the constant control of the level of odorants in a pipeline is required by law to prevent explosions and accidents. Currently, the detection of the most common odorants (THT = tetrahydrothiophane; TBM = tertiary butyl mercaptan) added to natural gas streams in pipelines is made in situ by using portable gas chromatography apparatuses. In this study, we report the analysis of the ultraviolet spectra obtained by a customized ultraviolet spectrophotometer, named Spectra, for the in-field detection of thiophane and tertiary butyl mercaptan. Spectra were conceived to accomplish the remote analysis of odorants in the pipelines of the natural gas stream through the adoption of technical solutions aimed to adapt a basic bench ultraviolet spectrophotometer to the in-field analysis of gases. The remotely controlled system acquires spectra continuously, performing the quantitative determination of odorants and catching systemic or accidental variations of the gaseous mixture in different sites of the pipeline. The analysis of the experimental spectra was carried out also through theoretical quantum mechanical approaches aimed to detect and to correctly assign the nature of the intrinsic electronic transitions of the two odorants, thiophane and tertiary butyl mercaptan, that cause the ultraviolet absorptions. So far, these theoretical aspects have never been studied before. The absorption maxima of thiophane and tertiary butyl mercaptan spectra were computationally simulated through the usage of selected molecular models with satisfactory results. The good matches between the experimental and theoretical datasets corroborate the reliability of the collected data. During the tests, unexpected pollutants and accidental malfunctions have been detected and also identified by Spectra, making this instrument suitable for many purposes.
2021
Istituto di Chimica dei Composti OrganoMetallici - ICCOM -
field detection
natural gas odorants
Portable ultraviolet spectrophotometer
TD-DFT
tetrahydrothiophane
THT detection
time-dependent density functional theory calculations
vibronic structure
File in questo prodotto:
File Dimensione Formato  
prod_434231-doc_175852.pdf

solo utenti autorizzati

Descrizione: Odorant Monitoring in Natural Gas Pipelines Using Ultraviolet-Visible Spectroscopy
Tipologia: Versione Editoriale (PDF)
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 648.85 kB
Formato Adobe PDF
648.85 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/379115
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? ND
social impact