The DC electronic response of a small area polycrystalline diamond film irradiated by a low energy X-ray beam has been analysed. The relationship between the induced photocurrent and the X-ray flux has been investigated making use of an X-ray absorber method. Aluminium foils have been used to attenuate the direct beam and the metaldiamondmetal device response tested at several electric field values. The X-ray photocurrent shows a linear response of the device that has been observed for a large range of X-ray flux values.
CVD diamond dosimetric response evaluated by X-ray absorbers method
E Cappelli;
2003
Abstract
The DC electronic response of a small area polycrystalline diamond film irradiated by a low energy X-ray beam has been analysed. The relationship between the induced photocurrent and the X-ray flux has been investigated making use of an X-ray absorber method. Aluminium foils have been used to attenuate the direct beam and the metaldiamondmetal device response tested at several electric field values. The X-ray photocurrent shows a linear response of the device that has been observed for a large range of X-ray flux values.File in questo prodotto: