In this work, the development and testing of a novel fiber-optic based label-free biosensor is presented, whose performance were verified through the detection of C-reactive protein (CRP) in serum. The device is based on a long period grating fabricated in a double cladding fiber with a W-shaped refractive index (RI) profile. As a result, the working point of the device was tuned to the mode transition region by chemical etching of the outer fiber cladding, obtaining a significant enhancement of the RI sensitivity and an excellent visibility of the grating resonances due to the mode transition in an all-silica structure. The fiber transducer was coated with a nanometric thin layer of graphene oxide in order to provide functional groups for the covalent immobilization of the biological recognition element. A very low limit of detection of about 0.15?ng/mL was obtained during the detection of CRP in serum, and a large working range (1?ng/mL - 100??g/mL) of clinical relevance has been also achieved.

Long period grating in double cladding fiber coated with graphene oxide as high-performance optical platform for biosensing

Lucia Sansone;Francesco Baldini;Francesco Chiavaioli
;
Michele Giordano
;
Ambra Giannetti;
2021

Abstract

In this work, the development and testing of a novel fiber-optic based label-free biosensor is presented, whose performance were verified through the detection of C-reactive protein (CRP) in serum. The device is based on a long period grating fabricated in a double cladding fiber with a W-shaped refractive index (RI) profile. As a result, the working point of the device was tuned to the mode transition region by chemical etching of the outer fiber cladding, obtaining a significant enhancement of the RI sensitivity and an excellent visibility of the grating resonances due to the mode transition in an all-silica structure. The fiber transducer was coated with a nanometric thin layer of graphene oxide in order to provide functional groups for the covalent immobilization of the biological recognition element. A very low limit of detection of about 0.15?ng/mL was obtained during the detection of CRP in serum, and a large working range (1?ng/mL - 100??g/mL) of clinical relevance has been also achieved.
2021
Istituto di Fisica Applicata - IFAC
Istituto per i Polimeri, Compositi e Biomateriali - IPCB - Sede Secondaria di Napoli (Portici)
Label-free biosensors
Real-time biomolecule detection
Double cladding fiber sensors
Graphene oxide
Long period gratings
C-Reactive protein
File in questo prodotto:
File Dimensione Formato  
prod_435255-doc_170048.pdf

solo utenti autorizzati

Descrizione: Long period grating in double cladding fiber coated with graphene oxide as high-performance optical platform for biosensing
Tipologia: Versione Editoriale (PDF)
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 5.75 MB
Formato Adobe PDF
5.75 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/383918
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 129
  • ???jsp.display-item.citation.isi??? 115
social impact