In view of high-precision molecular spectroscopy in the infrared, we realized a narrow-linewidth laser source at 5.8 mu m traceable to the second in the International System of Units by mixing two near-infrared lasers in an orientation-patterned GaP crystal. The generated radiation is traceable to a Cs fountain through an optical fiber-delivered carrier; its estimated linewidth is 2.3 kHz at 1 s integration time. A quantum cascade laser emitting at the same wavelength is used to boost the power of the mid-IR radiation up to few tens of mW. Thanks to a phase-locking loop, about 70% of the quantum cascade laser output power is within the linewidth of the difference-frequency radiation. This apparatus is used for absolute molecular spectroscopy in a region, above 5 mu m, where high-precision traceability and reliable operation are still challenging. In this paper, we provide the basic information to implement a powerful and metrological grade source in the mid-infrared for precision measurements.

A QCL-based metrological-grade source at 6 mu m

D'Ambrosio D
Primo
;
Borri S
;
De Natale P;Insero G;
2020

Abstract

In view of high-precision molecular spectroscopy in the infrared, we realized a narrow-linewidth laser source at 5.8 mu m traceable to the second in the International System of Units by mixing two near-infrared lasers in an orientation-patterned GaP crystal. The generated radiation is traceable to a Cs fountain through an optical fiber-delivered carrier; its estimated linewidth is 2.3 kHz at 1 s integration time. A quantum cascade laser emitting at the same wavelength is used to boost the power of the mid-IR radiation up to few tens of mW. Thanks to a phase-locking loop, about 70% of the quantum cascade laser output power is within the linewidth of the difference-frequency radiation. This apparatus is used for absolute molecular spectroscopy in a region, above 5 mu m, where high-precision traceability and reliable operation are still challenging. In this paper, we provide the basic information to implement a powerful and metrological grade source in the mid-infrared for precision measurements.
2020
Istituto Nazionale di Ottica - INO
laser stabilization
metrology
mid-infrared spectroscopy
molecular beams
File in questo prodotto:
File Dimensione Formato  
38 - D_Ambrosio - APB 126 - 2020 - A QCL-based metrological-grade source at 6 um.pdf

solo utenti autorizzati

Tipologia: Versione Editoriale (PDF)
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 1.98 MB
Formato Adobe PDF
1.98 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
APB-Dambrosio2019-Rev 2.pdf

accesso aperto

Tipologia: Documento in Post-print
Licenza: Altro tipo di licenza
Dimensione 3.15 MB
Formato Adobe PDF
3.15 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/387843
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact