Single-photon detectors and bolometers represent the bridge between different topics in science, such as quantum computation, astronomy, particle physics, and biology. Nowadays, superconducting bolometers and calorimeters are the most-sensitive detectors in the terahertz and subterahertz bands. Here, we propose and demonstrate a Josephson escape sensor (JES) that could find natural application in astrophysics. The JES is composed of a fully superconducting one-dimensional Josephson junction, whose resistance-versus-temperature characteristics can be precisely controlled by a bias current. Therefore, differently from traditional superconducting detectors, the JES sensitivity and working temperature can be in situ simply and finely tuned depending on the application requirements. A JES bolometer is expected to show an intrinsic thermal-fluctuation-noise noise-equivalent power on the order of 10(-25) W/Hz(1/2), while a JES calorimeter could provide a frequency resolution of about 2 GHz, as deduced from the experimental data. In addition, the sensor can operate at the critical temperature (i.e., working as a conventional transition-edge sensor), with a noise-equivalent power of approximately 6 x 10(-20) W/Hz(1/2) and a frequency resolution of approximately 100 GHz.

Hypersensitive Tunable Josephson Escape Sensor for Gigahertz Astronomy

Paolucci Federico;Ligato Nadia;Giazotto Francesco
2020

Abstract

Single-photon detectors and bolometers represent the bridge between different topics in science, such as quantum computation, astronomy, particle physics, and biology. Nowadays, superconducting bolometers and calorimeters are the most-sensitive detectors in the terahertz and subterahertz bands. Here, we propose and demonstrate a Josephson escape sensor (JES) that could find natural application in astrophysics. The JES is composed of a fully superconducting one-dimensional Josephson junction, whose resistance-versus-temperature characteristics can be precisely controlled by a bias current. Therefore, differently from traditional superconducting detectors, the JES sensitivity and working temperature can be in situ simply and finely tuned depending on the application requirements. A JES bolometer is expected to show an intrinsic thermal-fluctuation-noise noise-equivalent power on the order of 10(-25) W/Hz(1/2), while a JES calorimeter could provide a frequency resolution of about 2 GHz, as deduced from the experimental data. In addition, the sensor can operate at the critical temperature (i.e., working as a conventional transition-edge sensor), with a noise-equivalent power of approximately 6 x 10(-20) W/Hz(1/2) and a frequency resolution of approximately 100 GHz.
2020
Istituto Nanoscienze - NANO
Inglese
14
3
9
Sì, ma tipo non specificato
Ghz sensors
bolometer
calorimeter
Internazionale
6
info:eu-repo/semantics/article
262
Paolucci, Federico; Ligato, Nadia; Buccheri, Vittorio; Germanese, Gaia; Virtanen, Pauli; Giazotto, Francesco
01 Contributo su Rivista::01.01 Articolo in rivista
open
   breAkThrough innovaTion pRogrAmme for a pan-European Detection and Imaging eCosysTem
   ATTRACT
   European Commission
   Horizon 2020 Framework Programme
   777222

   Thermoelectric detector based on superconductor-ferromagnet heterostructures
   SUPERTED
   European Commission
   Horizon 2020 Framework Programme
   800923

   INFNRT2 172800
   Regione Toscana
   POR FSE 2014-2020
   INFNRT2 172800
File in questo prodotto:
File Dimensione Formato  
2003.05966v3.pdf

accesso aperto

Tipologia: Documento in Pre-print
Licenza: Altro tipo di licenza
Dimensione 1.27 MB
Formato Adobe PDF
1.27 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/391515
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 10
social impact