This paper presents a complete, compact, and low power consumption instrument designed for time-domain near-infrared spectroscopy. It employs two custom-designed pulsed diode lasers (operating at 830 and 670 nm, with average optical power higher than 2 mW at 40 MHz repetition frequency), a single-photon detection module (based on a 1 mm2 active area silicon photomultiplier), and a custom time-to-digital converter with 10 ps time resolution. The system experimental characterization shows an instrument response function narrower than 300 ps (full-width at half maximum), with measurement stability better than ?1% over several hours of operation. The instrument, which is housed into a compact aluminum case (size 200 ? 160 ? 50 mm3), is specifically tailored for portability and ease of operation, hence fostering the diffusion of time-domain diffuse optics techniques. Thanks to a total power consumption lower than 10 W, this system is suitable for battery operation, thus enabling on-field measurements.

A Compact Two-Wavelength Time-Domain NIRS System Based on SiPM and Pulsed Diode Lasers

A Torricelli;A Pifferi;
2017

Abstract

This paper presents a complete, compact, and low power consumption instrument designed for time-domain near-infrared spectroscopy. It employs two custom-designed pulsed diode lasers (operating at 830 and 670 nm, with average optical power higher than 2 mW at 40 MHz repetition frequency), a single-photon detection module (based on a 1 mm2 active area silicon photomultiplier), and a custom time-to-digital converter with 10 ps time resolution. The system experimental characterization shows an instrument response function narrower than 300 ps (full-width at half maximum), with measurement stability better than ?1% over several hours of operation. The instrument, which is housed into a compact aluminum case (size 200 ? 160 ? 50 mm3), is specifically tailored for portability and ease of operation, hence fostering the diffusion of time-domain diffuse optics techniques. Thanks to a total power consumption lower than 10 W, this system is suitable for battery operation, thus enabling on-field measurements.
2017
Istituto di fotonica e nanotecnologie - IFN
Diode lasers
Instruments;Semiconductor lasers;Time-domain analysis;Measurement by laser beam;Optics;Photonics;Time-domain;diffuse optics;near-infrared spectroscopy;single-photon;silicon photomultiplier;pulsed diode laser;time-to-digital converter
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/401415
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 51
  • ???jsp.display-item.citation.isi??? ND
social impact