This paper presents a parametric automatic procedure to calibrate the multichannel Rayleigh-Mie-Raman lidar at the Institute for Atmospheric Science and Climate of the Italian National Research Council (ISAC-CNR) in Tor Vergata, Rome, Italy, using as a reference the operational 0000 UTC soundings at the WMO station 16245 (Pratica di Mare) located about 25 km southwest of the lidar site. The procedure, which is applied to both channels of the system, first identifies portions of the lidar and radiosonde profiles that are assumed to sample the same features of the water vapor profile, taking into account the different time and space sampling. Then, it computes the calibration coefficient with a best-fit procedure, weighted by the instrumental errors of both radiosounding and lidar. The parameters to be set in the procedure are described, and values adopted are discussed. The procedure was applied to a set of 57 sessions of nighttime 1-min-sampling lidar profiles (roughly about 300 h of measurements) covering the whole annual cycle (February 2007-September 2008). A calibration coefficient is computed for each measurement session. The variability of the calibration coefficients (similar to 10%) over periods with the same instrumental setting is reduced compared to the values obtained with the previously adopted, operator-assisted, and time-consuming calibration procedure. Reduction of variability, as well as the absence of evident trends, gives confidence both on system stability as well as on the developed procedure. Because of the definition of the calibration coefficient and of the different sampling between lidar and radiosonde, a contribution to the variability resulting from aerosol extinction and to the spatial and temporal variability of the water vapor mixing ratio is expected. A preliminary analysis aimed at identifying the contribution to the variability from these factors is presented. The parametric nature of the procedure makes it suitable for application to similar Raman lidar systems.

Calibration of a Multichannel Water Vapor Raman Lidar through Noncollocated Operational Soundings: Optimization and Characterization of Accuracy and Variability

Dionisi D;F Congeduti;G Liberti;F Cardillo
2010

Abstract

This paper presents a parametric automatic procedure to calibrate the multichannel Rayleigh-Mie-Raman lidar at the Institute for Atmospheric Science and Climate of the Italian National Research Council (ISAC-CNR) in Tor Vergata, Rome, Italy, using as a reference the operational 0000 UTC soundings at the WMO station 16245 (Pratica di Mare) located about 25 km southwest of the lidar site. The procedure, which is applied to both channels of the system, first identifies portions of the lidar and radiosonde profiles that are assumed to sample the same features of the water vapor profile, taking into account the different time and space sampling. Then, it computes the calibration coefficient with a best-fit procedure, weighted by the instrumental errors of both radiosounding and lidar. The parameters to be set in the procedure are described, and values adopted are discussed. The procedure was applied to a set of 57 sessions of nighttime 1-min-sampling lidar profiles (roughly about 300 h of measurements) covering the whole annual cycle (February 2007-September 2008). A calibration coefficient is computed for each measurement session. The variability of the calibration coefficients (similar to 10%) over periods with the same instrumental setting is reduced compared to the values obtained with the previously adopted, operator-assisted, and time-consuming calibration procedure. Reduction of variability, as well as the absence of evident trends, gives confidence both on system stability as well as on the developed procedure. Because of the definition of the calibration coefficient and of the different sampling between lidar and radiosonde, a contribution to the variability resulting from aerosol extinction and to the spatial and temporal variability of the water vapor mixing ratio is expected. A preliminary analysis aimed at identifying the contribution to the variability from these factors is presented. The parametric nature of the procedure makes it suitable for application to similar Raman lidar systems.
2010
Istituto di Scienze dell'Atmosfera e del Clima - ISAC
LIDAR RAMAN
WATER VAPOR
CALIBRATION
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/41247
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 23
  • ???jsp.display-item.citation.isi??? ND
social impact