This paper describes a compact electronic system employing a synchronous demodulation measurement method for the acquisition of pulsed-current signals. The fabricated prototype shows superior performance in terms of signal-to-noise ratio in comparison to conventional instrumentation performing free-running measurements, especially when extremely narrow pulses are concerned. It shows a reading error around 0.1% independently of the signal duty cycle (D) in the investigated D = 10-4-10-3 range. Conversely, high-precision electrometers display reading errors as high as 30% for a D = 10-4, which reduces to less than 1% only for D > 3 × 10-3. Field tests demonstrate that the developed front-end/readout electronics is particularly effective when coupled to dosimeters irradiated with the X-rays sourced by a medical linear accelerator. Therefore, it may surely be exploited for the real-time monitoring of the dosimeter output current, as required in modern radiotherapy techniques employing ultra-narrow pulses of high-energy photons or nuclear particles.
Accurate Signal Conditioning for Pulsed-Current Synchronous Measurements
Sara Pettinato;Marco Girolami;Stefano Salvatori
2022
Abstract
This paper describes a compact electronic system employing a synchronous demodulation measurement method for the acquisition of pulsed-current signals. The fabricated prototype shows superior performance in terms of signal-to-noise ratio in comparison to conventional instrumentation performing free-running measurements, especially when extremely narrow pulses are concerned. It shows a reading error around 0.1% independently of the signal duty cycle (D) in the investigated D = 10-4-10-3 range. Conversely, high-precision electrometers display reading errors as high as 30% for a D = 10-4, which reduces to less than 1% only for D > 3 × 10-3. Field tests demonstrate that the developed front-end/readout electronics is particularly effective when coupled to dosimeters irradiated with the X-rays sourced by a medical linear accelerator. Therefore, it may surely be exploited for the real-time monitoring of the dosimeter output current, as required in modern radiotherapy techniques employing ultra-narrow pulses of high-energy photons or nuclear particles.| File | Dimensione | Formato | |
|---|---|---|---|
|
prod_469371-doc_190071.pdf
accesso aperto
Descrizione: Sensors 2022
Tipologia:
Versione Editoriale (PDF)
Licenza:
Creative commons
Dimensione
2.27 MB
Formato
Adobe PDF
|
2.27 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.