The study aim was to evaluate and compare the analytical performance of the new chemiluminescent immunoassay for cardiac troponin I (cTnI), called Access hs-TnI using DxI platform, with those of Access AccuTnI+3 method, and high-sensitivity (hs) cTnI method for ARCHITECT platform. The limits of blank (LoB), detection (LoD) and quantitation (LoQ) at 10% and 20% CV were evaluated according to international standardized protocols. For the evaluation of analytical performance and comparison of cTnI results, both heparinized plasma samples, collected from healthy subjects and patients with cardiac diseases, and quality control samples distributed in external quality assessment programs were used. LoB, LoD and LoQ at 20% and 10% CV values of the Access hs-cTnI method were 0.6, 1.3, 2.1 and 5.3 ng/L, respectively. Access hs-cTnI method showed analytical performance significantly better than that of Access AccuTnI+3 method and similar results to those of hs ARCHITECT cTnI method. Moreover, the cTnI concentrations measured with Access hs-cTnI method showed close linear regressions with both Access AccuTnI+3 and ARCHITECT hs-cTnI methods, although there were systematic differences between these methods. There was no difference between cTnI values measured by Access hs-cTnI in heparinized plasma and serum samples, whereas there was a significant difference between cTnI values, respectively measured in EDTA and heparin plasma samples. Access hs-cTnI has analytical sensitivity parameters significantly improved compared to Access AccuTnI+3 method and is similar to those of the high-sensitivity method using ARCHITECT platform.

Evaluation of analytical performance of a new high-sensitivity immunoassay for cardiac troponin i

Ndreu R;
2018

Abstract

The study aim was to evaluate and compare the analytical performance of the new chemiluminescent immunoassay for cardiac troponin I (cTnI), called Access hs-TnI using DxI platform, with those of Access AccuTnI+3 method, and high-sensitivity (hs) cTnI method for ARCHITECT platform. The limits of blank (LoB), detection (LoD) and quantitation (LoQ) at 10% and 20% CV were evaluated according to international standardized protocols. For the evaluation of analytical performance and comparison of cTnI results, both heparinized plasma samples, collected from healthy subjects and patients with cardiac diseases, and quality control samples distributed in external quality assessment programs were used. LoB, LoD and LoQ at 20% and 10% CV values of the Access hs-cTnI method were 0.6, 1.3, 2.1 and 5.3 ng/L, respectively. Access hs-cTnI method showed analytical performance significantly better than that of Access AccuTnI+3 method and similar results to those of hs ARCHITECT cTnI method. Moreover, the cTnI concentrations measured with Access hs-cTnI method showed close linear regressions with both Access AccuTnI+3 and ARCHITECT hs-cTnI methods, although there were systematic differences between these methods. There was no difference between cTnI values measured by Access hs-cTnI in heparinized plasma and serum samples, whereas there was a significant difference between cTnI values, respectively measured in EDTA and heparin plasma samples. Access hs-cTnI has analytical sensitivity parameters significantly improved compared to Access AccuTnI+3 method and is similar to those of the high-sensitivity method using ARCHITECT platform.
2018
Istituto di Fisiologia Clinica - IFC
analytical sensitivity
automated immunometric assays
cardiac biomarkers
cardiac troponins
high-sensitivity methods
methods comparison
myocardial infarction
quality specifications
quality control materials
File in questo prodotto:
File Dimensione Formato  
prod_446269-doc_160451.pdf

solo utenti autorizzati

Descrizione: Ricerca cardiovascolare
Tipologia: Versione Editoriale (PDF)
Dimensione 515.41 kB
Formato Adobe PDF
515.41 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/424851
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 52
  • ???jsp.display-item.citation.isi??? ND
social impact