We propose a new quantitative electrochemical method for determining glucose-6-phosphate dehydrogenase (EC 1.1.1.49) activity in purified erythrocytes or in whole blood, based on measurement of the pH change caused by oxidation of glucose 6-phosphate to 6-phosphogluconic acid, with simultaneous reduction of NADP+ to NADPH + H+. No sample pretreatment (e.g., preparation of hemolysate) is needed, and the automatic correction for sample blanks obviates interference from 6-phosphogluconate dehydrogenase (EC 1.1.1.44). The method is simple and fast, and the standard curve is linear to at least 2200 U/L at 37 degrees C. The within-day CV was 3.9% for activities in healthy individuals (mean value 1204 U/L), and 10% for deficient ones (classified as belonging to class II, mean value 407 U/L). Results (y) correlated well with those obtained with the WHO-recommended method (x): y = 1.13x + 0.02 (r = 0.971) for purified erythrocytes. Normal reference intervals are reported for the enzyme in purified erythrocytes and in whole blood.

Erythrocytic glucose-6-phosphatedehydrogenase measured by a differential pH technique.

Ripamonti M;
1987

Abstract

We propose a new quantitative electrochemical method for determining glucose-6-phosphate dehydrogenase (EC 1.1.1.49) activity in purified erythrocytes or in whole blood, based on measurement of the pH change caused by oxidation of glucose 6-phosphate to 6-phosphogluconic acid, with simultaneous reduction of NADP+ to NADPH + H+. No sample pretreatment (e.g., preparation of hemolysate) is needed, and the automatic correction for sample blanks obviates interference from 6-phosphogluconate dehydrogenase (EC 1.1.1.44). The method is simple and fast, and the standard curve is linear to at least 2200 U/L at 37 degrees C. The within-day CV was 3.9% for activities in healthy individuals (mean value 1204 U/L), and 10% for deficient ones (classified as belonging to class II, mean value 407 U/L). Results (y) correlated well with those obtained with the WHO-recommended method (x): y = 1.13x + 0.02 (r = 0.971) for purified erythrocytes. Normal reference intervals are reported for the enzyme in purified erythrocytes and in whole blood.
1987
glucose-6-phosphate dehydrogenase
whole blood
differential pH technique.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/432462
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact