The application of a new technique, based on differential measurements of pH, to determine urea concentration in patients of a dialysis center, is reported. Urea in plasma, whole blood or dialysis fluids is measured by an enzymatic reaction, with urease; the procedure, requiring 10 microL of sample, is simple, fast and correlates well with a reference spectrophotometric method, in the 0-300 mg/dL concentration range, according to the equation y = 1.0291 X -0.0777; r = 0.9991; n = 73.

UREA DETERMINATION IN DIALYSIS, BASED ON A DIFFERENTIAL PH TECHNIQUE

RIPAMONTI M;
1987

Abstract

The application of a new technique, based on differential measurements of pH, to determine urea concentration in patients of a dialysis center, is reported. Urea in plasma, whole blood or dialysis fluids is measured by an enzymatic reaction, with urease; the procedure, requiring 10 microL of sample, is simple, fast and correlates well with a reference spectrophotometric method, in the 0-300 mg/dL concentration range, according to the equation y = 1.0291 X -0.0777; r = 0.9991; n = 73.
1987
urea
differential measurements of pH
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/432463
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 3
social impact