We have developed a method that needs only the computed tomography (CT) indicator box to coregister positron emission tomography (PET) images and integrates this information with magnetic resonance imaging. The study was performed using a PET/CT scanner. A standard CT bed adapter was attached to the scanner couch. Then, the patient, with the Leksell G frame fixed, was positioned into the scanner with the CT indicator box. PET images were acquired using either [18F]fluorodeoxyglucose or [11C]choline as radioisotopes. After acquisition, CT and PET images were exported in DICOM 3 standard and transferred to a dedicated workstation via data link. Homemade software was implemented for multimodal image fusion. PET images were overwritten to their corresponding CT point values using a threshold algorithm, maintaining the stereotactic CT markers. The use of a CT indicator simplifies the procedure, because there is no need for a radioactive solution filling the indicator box. This method was tested first using a phantom and then in patients. The localization accuracy of the PET images is limited only by the slice thickness.

A simplified method to integrate metabolic images in stereotactic procedures using a PET-CT scanner

Rizzo G;Fazio F
2005

Abstract

We have developed a method that needs only the computed tomography (CT) indicator box to coregister positron emission tomography (PET) images and integrates this information with magnetic resonance imaging. The study was performed using a PET/CT scanner. A standard CT bed adapter was attached to the scanner couch. Then, the patient, with the Leksell G frame fixed, was positioned into the scanner with the CT indicator box. PET images were acquired using either [18F]fluorodeoxyglucose or [11C]choline as radioisotopes. After acquisition, CT and PET images were exported in DICOM 3 standard and transferred to a dedicated workstation via data link. Homemade software was implemented for multimodal image fusion. PET images were overwritten to their corresponding CT point values using a threshold algorithm, maintaining the stereotactic CT markers. The use of a CT indicator simplifies the procedure, because there is no need for a radioactive solution filling the indicator box. This method was tested first using a phantom and then in patients. The localization accuracy of the PET images is limited only by the slice thickness.
2005
Istituto di Bioimmagini e Fisiologia Molecolare - IBFM
CT/PET
software GammaPlan,
radiochirurgia con Gamma Knife,
Elaborazione di immagini multimodali
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/452474
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact