Experimental results obtained with commercially available PSPMs coupled to 0.5-mm diameter scintillating fiber arrays show space resolution of < 200 ?m and time resolution better than 1 ns with a detection efficiency of about 90%. In addition, the preliminary studies indicate the feasibility of separating double-tracks with a minimum distance of 3 mm using this technology.

FAST READOUT OF SCINTILLATING-FIBER ARRAYS USING POSITION-SENSITIVE PHOTOMULTIPLIERS

G Salvato;
1995

Abstract

Experimental results obtained with commercially available PSPMs coupled to 0.5-mm diameter scintillating fiber arrays show space resolution of < 200 ?m and time resolution better than 1 ns with a detection efficiency of about 90%. In addition, the preliminary studies indicate the feasibility of separating double-tracks with a minimum distance of 3 mm using this technology.
1995
Istituto per i Processi Chimico-Fisici - IPCF
Inglese
357
1
78
86
9
http://dx.doi.org/10.1016/0168-9002(94)01399-3
Sì, ma tipo non specificato
HODOSCOPE
1
info:eu-repo/semantics/article
262
V Agoritsas; N Akchurin; AM Bergdolt; A Bravar; O Bing; J Ditta; J Dufournaud; R Giacomich; AM Gorin; K Kuroda; D Magaudda; IV Manuilov; A Michalowicz...espandi
01 Contributo su Rivista::01.01 Articolo in rivista
none
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/455086
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 10
social impact