Two compact generators of 14 and 2.5 MeV neutrons have been installed in the new Neutron Irradiation Laboratory for Electronics (NILE) facility. Neutrons are produced by deuterium-deuterium (DD) and deuterium-tritium (DT) fusion reactions. The facility has been designed with the assistance of Monte Carlo calculations. Here, we present neutron measurements with multiple techniques for the characterization of the radiation field and dosimetry. Activation foils are used as a standard technique for passive measurements of neutron fluxes. Active detectors, like fission chambers, are used mainly as online neutron monitor. Diamond detectors are applied with the main purpose of high-resolution neutron spectroscopy. Scintillators are used as neutron monitors and spectrometers with the additional capability of neutron/gamma discrimination. The characterized neutron beams will be mainly applied for the testing of electronics.

Characterization Measurements of Compact Neutron Generators of the New NILE Facility

Tardocchi M;Perelli Cippo E;Rigamonti D;
2023

Abstract

Two compact generators of 14 and 2.5 MeV neutrons have been installed in the new Neutron Irradiation Laboratory for Electronics (NILE) facility. Neutrons are produced by deuterium-deuterium (DD) and deuterium-tritium (DT) fusion reactions. The facility has been designed with the assistance of Monte Carlo calculations. Here, we present neutron measurements with multiple techniques for the characterization of the radiation field and dosimetry. Activation foils are used as a standard technique for passive measurements of neutron fluxes. Active detectors, like fission chambers, are used mainly as online neutron monitor. Diamond detectors are applied with the main purpose of high-resolution neutron spectroscopy. Scintillators are used as neutron monitors and spectrometers with the additional capability of neutron/gamma discrimination. The characterized neutron beams will be mainly applied for the testing of electronics.
2023
Istituto per la Scienza e Tecnologia dei Plasmi - ISTP
Activation foils
diamond detectors
dosimetry
neutron generators
scintillators
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/464449
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? ND
social impact