The trackability of tiny endovascular medical implants (TEMIs) during and after surgery is a crucial parameter for their accurate implantation and after surgery follow-ups. In this work, novel radiopaque Ta-W coatings were produced by magnetron sputtering deposition and further investigated with the aim to improve X-ray visibility of TEMIs under fluoroscopy imaging. Different deposition strategies, named micro-multilayer (Micro-ML), nano-multilayer (Nano-ML) and co-deposition (Co-dep), and different substrate temperatures (25 °C and 600 °C) were implemented which allowed to tune the morphology, mechanical properties and radiopacity of the coatings, while avoiding the formation of brittle phases, such as β-Ta and β-W. Ta-W coatings produced at high temperature (600 °C) contained stable phases such as α-Ta (bcc) and α-W (bcc), without brittle phases such as β-Ta and β-W. However, traces of β-W and β-Ta were detected in the group of samples produced at low substrate temperature (25 °C). The elastic modulus and the hardness of Micro-ML and Nano-ML were found to increase with increasing temperature, whereas they decreased in the case of samples produced by the Co-dep condition. The chemical state of Ta as well as the proportion of oxidized Ta (Ta-O) detected on all coatings were found to change as a function of the substrate temperature and deposition strategy. Samples with the highest amount of Ta-O at their surfaces showed the lowest corrosion potential. Finally, the samples produced by Micro-ML strategies performed on 600 °C substrates and irradiated according to radiographic imaging of conditions comparable to that of fluoroscopy, exhibited an X-ray contrast improvement of 82%/μm, which was higher than what was found in all other samples. Higher volume fractions of α-W in the coatings resulted in higher X-ray attenuation potential for photons in the energy range typical of fluoroscopy. Overall, the production of Ta-W coatings by the Micro-ML strategy and performed with a substrate temperature of 600 °C seems an optimal approach for the production of highly visible TEMIs in X-ray imaging.

The impact of structure and temperature on the mechanical properties and radiopacity of Ta-W coatings for tiny endovascular medical implants

Sofia Gambaro;
2023

Abstract

The trackability of tiny endovascular medical implants (TEMIs) during and after surgery is a crucial parameter for their accurate implantation and after surgery follow-ups. In this work, novel radiopaque Ta-W coatings were produced by magnetron sputtering deposition and further investigated with the aim to improve X-ray visibility of TEMIs under fluoroscopy imaging. Different deposition strategies, named micro-multilayer (Micro-ML), nano-multilayer (Nano-ML) and co-deposition (Co-dep), and different substrate temperatures (25 °C and 600 °C) were implemented which allowed to tune the morphology, mechanical properties and radiopacity of the coatings, while avoiding the formation of brittle phases, such as β-Ta and β-W. Ta-W coatings produced at high temperature (600 °C) contained stable phases such as α-Ta (bcc) and α-W (bcc), without brittle phases such as β-Ta and β-W. However, traces of β-W and β-Ta were detected in the group of samples produced at low substrate temperature (25 °C). The elastic modulus and the hardness of Micro-ML and Nano-ML were found to increase with increasing temperature, whereas they decreased in the case of samples produced by the Co-dep condition. The chemical state of Ta as well as the proportion of oxidized Ta (Ta-O) detected on all coatings were found to change as a function of the substrate temperature and deposition strategy. Samples with the highest amount of Ta-O at their surfaces showed the lowest corrosion potential. Finally, the samples produced by Micro-ML strategies performed on 600 °C substrates and irradiated according to radiographic imaging of conditions comparable to that of fluoroscopy, exhibited an X-ray contrast improvement of 82%/μm, which was higher than what was found in all other samples. Higher volume fractions of α-W in the coatings resulted in higher X-ray attenuation potential for photons in the energy range typical of fluoroscopy. Overall, the production of Ta-W coatings by the Micro-ML strategy and performed with a substrate temperature of 600 °C seems an optimal approach for the production of highly visible TEMIs in X-ray imaging.
2023
Istituto di Chimica della Materia Condensata e di Tecnologie per l'Energia (ICMATE) - Sede Secondaria Genova
Magnetron sputtering
Ta-W multilayer coatings
Thin films
Radiopacity measurement
Fluoroscopy
Tiny endovascular medical implants
File in questo prodotto:
File Dimensione Formato  
The impact of structure and temperature on the mechanical properties and radiopacity of Ta.pdf

accesso aperto

Tipologia: Documento in Pre-print
Licenza: Creative commons
Dimensione 3 MB
Formato Adobe PDF
3 MB Adobe PDF Visualizza/Apri
Gambaro et al_Elsevier-2023.pdf

solo utenti autorizzati

Descrizione: Versione editoriale
Tipologia: Versione Editoriale (PDF)
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 923.07 kB
Formato Adobe PDF
923.07 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/534503
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact