Remotely sensed data, recorded by means of the MIVIS hyperspectral scanner in the framework of the research activity of the CNR Institutes IIA-LARA and IMAA, have been calibrated to reflectance values and then quantitatively compared with ground data. A new procedure for radiometric calibration has been defined by utilizing the MIVIS test-bench and applying a wider radiance range with respect to the one provided by the manufacturing company. New calibration curves have been determined and applied in the pre-processing chain. For validation purpose ground spectra were measured during the campaign by means of a portable spectroradiometer. The atmospheric correction has been carried out by implementing an IDL procedure to manage MODTRAN4 input and output cards. MIVIS test data acquired over Passo Corese (Roma) have shown how the new calibration coefficients significantly improve the radiometric accuracy. In particular, in the VIS spectral region the percentage error, with respect to a ground truth spectrum, is about half of that occurring if the standard calibration coefficients are used.

Laboratory activity for a new procedure of MIVIS calibration and relative validation with test data

Bassani, C;Cavalli, ROSA MARIA;Madonna, F;
2006

Abstract

Remotely sensed data, recorded by means of the MIVIS hyperspectral scanner in the framework of the research activity of the CNR Institutes IIA-LARA and IMAA, have been calibrated to reflectance values and then quantitatively compared with ground data. A new procedure for radiometric calibration has been defined by utilizing the MIVIS test-bench and applying a wider radiance range with respect to the one provided by the manufacturing company. New calibration curves have been determined and applied in the pre-processing chain. For validation purpose ground spectra were measured during the campaign by means of a portable spectroradiometer. The atmospheric correction has been carried out by implementing an IDL procedure to manage MODTRAN4 input and output cards. MIVIS test data acquired over Passo Corese (Roma) have shown how the new calibration coefficients significantly improve the radiometric accuracy. In particular, in the VIS spectral region the percentage error, with respect to a ground truth spectrum, is about half of that occurring if the standard calibration coefficients are used.
2006
Istituto di Metodologie per l'Analisi Ambientale - IMAA
Istituto sull'Inquinamento Atmosferico - IIA
radiometric calibrationMIVISatmospheric correctionMODTRAN
File in questo prodotto:
File Dimensione Formato  
admin-ingv,+06 Bassani.pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 821.75 kB
Formato Adobe PDF
821.75 kB Adobe PDF Visualizza/Apri